Robust Vacuum gauge for rough and fine vacuum range
Rugged combination sensor combines most advanced Pirani and capacitance membrane technology
High accuracy and excellent resolution in the range atmosphere to 5x10-5 mbar
Standard or chemical resistant sensor version with built-in LCD Display
Available as a package with all accessories for fine vacuum measurement
Multi-range vacuum gauge VMpro 1
DN16KF, 100-240V 50/60Hz
VMpro 1 multi-range vacuum gauge combines Pirani technology with the advantages of a ceramic capacitance diaphragm sensor in a single gauge.
In the measurement range between 10 mbar and atmospheric pressure the capacitance diaphragm technology provides gas-type independent, highly accurate values for reliable pressure measurement.
The built-in digital LCD display gives clear readability for easy vacuum level monitoring and also shows the current measuring principle. There is a status indication light on top of the sensor
Scope of delivery: VMpro 1 vacuum gauge, AC/DC power adapter with EU and UK plug adapters, connection cable
Multi-range vacuum gauge VMpro 2
Chemical resistant, DN16KF, 100-240V 50/60Hz
VMpro 2 chemical resistant multi-range vacuum gauge combines Pirani technology with the advantages of a ceramic capacitance diaphragm sensor in a single gauge.
In the measurement range between 10 mbar and atmospheric pressure the capacitance diaphragm technology provides gas-type independent, highly accurate values for reliable pressure measurement.
A Nickel Pirani filament is used to give good compatibility with a wide range of chemicals.
The built-in digital display gives clear readability for easy vacuum level monitoring and also shows the current measuring principle. There is a status indication light on top of the sensor.
Scope of delivery: VMpro 2 Chem vacuum gauge, AC/DC power adapter with EU and UK plug adapters, connection cable
Technical Data
Parameter
Data
VMpro 1 / VMpro 2
Measuring range [mbar]
1500 â 5x10-5
Measuring range [Torr]
1125 â 3.8x10-5
ACCURACY [% of reading]
5x10-4 â 1x10-3 mbar
±50
1x10-3 â 100 mbar
±15
100 â 950 mbar
±5
950 â 1050 mbar
±2.5
Repeatability [% of reading]
1x10-3 â 1100 mbar
±2
Admissible temperature [OC]
Operation
+10 â +50
Storage
-20 â +65
Bakeout at flange
≤ +80
Bakeout at flange long tube
≤ +250
Max Pressure [bar]
10
Vacuum connection [DN]
16 KF
Maximum altitude [m]
2000
Relative humidity, annual mean [%] (non-condensing)